Customization: | Available |
---|---|
After-sales Service: | 1 Year |
Warranty: | 1 Year |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
1,Light source: quartz halogen lamp
2,Detection path: Single optical path detection, two reference channels (blank reference channel, background reference channel)*
3, wavelength range: 340-850nm *
4,Filters: 8-bit filter wheel, standard three filters: 405, 450, 620nm.
5. Reading range: 0-6 Abs*.
6,Accuracy (405nm): ±1% (0-3Abs), ±2% (3-4Abs)
7,Accuracy (405nm): CV ≤ 0.2% (0 - 3 Abs), CV ≤ 1.0% (3 - 4 Abs), standard measurement mode
8,Resolution:0.001Abs
9,Linear range: 0-3Abs, 96-well plate, fast measurement mode; 0-4Abs, 96-well plate, standard measurement mode
10,Measurement speed: 6 s, 96-well plate, rapid measurement mode; 12 s, 96-well plate, standard measurement mode
11,Applicable plate type: 96-well plate
12,Oscillator: linear oscillation, three speeds can be selected
13,Key display: high-resolution color LCD display, graphical interface, support multi-language *
14, data storage: the instrument can store 100 measurement procedures and 100 sets of measurement results, data can be exported through the USB interface, with three USB ports are connected to a computer, U disk, printer *
15, software: including SkanIt software, installation is not limited