TOP insulation transformer oil breakdown voltage tester series IIJ-II-80kv
Brief introduction:
IIJ-II series insulating oils dielectric strength automatic tester is a new generation full automatization bdv testing instrument developed on the basis of the hcj-9905a model, which is designed according to the international standard iec-156 and national standard gb507-86 (insulating oils- determination of the dielectric strength) requirements, and meet the various users'feedback and opinions.
Using setting firstly, start up testing secondly method, it is adopted scm as the leading, the whole testing procedure automatically run and controlled by the microcomputer. It is simply operated and convenient used.
Application and features:
It is suitable for testing various kinds of insulating oils dielectric strength. The main features as the following:
1. It has the automatic detection function, ex. It could automatic turn into the reset mode and execute the voltage regulator return to zero.
2. It is adopted the mini type tpu-a panel printer to automatically print the testing results.
3. It could be altered the testing frequency, stirring and stewing time, voice controlled and light controlled alarm, continuous printing and none printing according to the user's requirements.
4. It is used the automatic magnetism oscillator stirring the insulating oil in the oil sample cup avoiding the oil sample misproportion and gas bubble.
Technical specifications
1. Power supply: Ac 220v +/-10%, 50 hz +/-5%
2. Measuring range: Ac 0-80 kv
3. Limited current: 5 ma
4. Measuring accuracy: 1%
5. Voltage regulating rate: 2kv/s +/-10%
6. Reservation setting:
Frequency: 1-9
Stirring time: 0-99 seconds
Stewing time: 0-999 seconds
7. Working condition:
Ambient temperature: 0 c-35 c
Relative humidity: ≤ 75%
8. Oil cup gap: 2.5 mm (the diameter of oil cup gapgauge)
9. Dimension: 415x315x315mm
10. N. W.: 28 kg
11. G. W.: 38 kg
Accessaries:
1. Operation instruction of oil tester 1 pc
2. Operation instruction of printer 1 pc
3. Special oil cup 1 pc
4. Power cable 1 pc
5.3A fuse 4 pc
6. Magnetic rotator 1 pc
Our oil tester/ analyzer: 1. Transformer oil Breakdown voltage Tester (BDV test kit) series IIJ-II
2. Acidity tester model ACD-3000I
3. Karl Fischer Water content Tester Model TP-6A / model TP-2100
4. Fully automatic Interfacial Tension Analyzer model IT-800
5. Gas-Chromatography transformer oil gas analyzer model DGA2013-1
6. Multi-functional Vibration tester model TYD-2000
7. Consolidation Point Determinator model PT-2000
8. Kinematic Viscosity tester model VST-8 /model VST-2000
9. Flash point tester (open cup) model TPO-3000
10. Flash point tester (close cup) model TPC-3000
11. TAN-DELTA TESTER, 10kV Capacitance and Dissipation Factor Test Set model CDEF
12. Automatic Freezing Point & Pour Point Tester model TP-501Z
13. Determination of foaming characteristic of lubricating oil model SD-12579
14. Earth resistance tester model HYDW-II
15. Coil Resistance Tester (TPZC-40)
16. CT/PT tester model TPVA-404
17. Transformer turn ratio tester model HYG-II
18. Relaying Protection Tester model TPJB-III
19. Digital Grounding Resistance Tester model TP2571-II
20. DC High Voltage Generator model ZGF
21. Underground water/gold detector model TP-UWD
Keywords: Oil purifier, oil purification, oil filtration, oil filter, oil purifying, oil filtrator, oil regeneration, oil recovery, oil tester, oil recycling, test equipment, oil water separator, oil reusing, waste oil disposal, energy saving
If you would like to know more about our oil purifiers and oil testers, please feel free to contact us any time.
Power supply |
AC220V±10%, 50Hz |
Output voltage |
0-80KV (60KV.80KV) |
Capacity |
1.2KVA (1.6KVA.2.0KVA) |
Speed of pressure rise |
about 2KV/S |
Voltage test precision |
±3% |
Breakdown sensitivity |
<2KV |
Wave difference |
<3% |
Time of breakdown |
<10ms |
Working altitude |
<1500m (if over 1500m can be designed specially) |
Dimension |
480mm*460mm*460mm |
Working condition |
Temperature |
30Celsius degree--40Celsius degree |
humidity |
The biggest comparatively humidity85% |
Reserve circumstance |
Temperature |
-20Celsius degree-60Celsius degree |
humidity |
The biggest comparatively humidity75% |