ISO Abel Closed Cup Flash Point Tester For Oils(TP-002C)
Application:
Applicable standard:GB/T 21789 ,ISO13736 , IP 170
Applicable range:Suitable for the determination of flash point of petroleum products and other liquids between -30 ° C and 70 ° C
Main Features:
Digital display heating voltage, intuitive and clear, easy to control heating process
Heating uses a solid state regulator to control the rate of temperature rise.
Electric ignition, eliminating the human error of the sweep time.
The test cup is covered with a stirring device, which can be freely disassembled for easytest
Stirring rate continuously adjustable to meet the needs of different samples.
Technical data
Test range: |
-30~70ºC(-30 ~ 18.5 °C need to be connected to a low temperature circulating cold bath) |
Flame diameter |
3~4mm |
Heating rate |
1ºC/min
|
Ambient temperature |
10ºC~30ºC |
Working environment |
No dust and corrosive gases, no strong magnetic field interference, inside the fume hood |
Total Power : |
500W |
Power: |
AC220V±10% |
Dimension |
390 mm×270 mm×430 mm
|
Weight: |
About 12 kg |
Work description (use in the range of 19 to 70 ° C)
First, fill the heating bath from the water injection port on the inclined surface of the instrument (this water should be about 10 °C lower than the room temperature), and add the test cup bath in the middle of the heating bath to the water with a depth of not less than 38 mm.
Take the sample into the test cup, cover the test cup cover, push the slide on the lid into the pull rod, and put the stirring upper sleeve into the lower sleeve to connect the gas hose.
Plug in the bath thermometer and sample thermometer.
After checking, connect the 220V power supply and turn on the power switch.
Adjust the heating potentiometer so that the sample in the test cup is heated at a rate of 1 ° C / min, and simultaneously turn on the stirring switch to adjust to the appropriate speed.
When the temperature of the test sample reaches 9 °C below the expected flash point, press the ignition switch to perform the ignition test.
An ignition test was performed for every 0.5 °C increase.
The temperature at which the flash point occurs is recorded when a flash point occurs.
At the end of the test, turn off the power. Clean the test cup and top cover to prepare for the next test.
About us:
Hello Friend!
Chongqing TOP is a professional leading manufacturer and supplier of oil purifiers and various related testers. We consists of departments respectively for technological development, machinery manufacture, instrument manufacture, quality control etc. On basis of mature development and manufacture technology and highly qualified production management team, the products own high quality, easy operation, excellent performance and lone service life.
Customer satisfaction is our utmost priority thus, customer service team is always on the brink to help our customers with their queries at any time.
We believe in working as a unit with an attitude of complete trust and understanding and are determined to provide safe, reliable and competitively priced products to our customers.
TOP Oil Purifiers uses single or multi-stage filtration to remove, particulate, free water, and small quantities of dissolved water if necessary. The systems are usually customized to suite the specific customer application. Oil "regeneration", "reclamation", "recycling" are all terms used to describe the process of restoring used, aged, contaminated oils to "like new" reusable condition. The reclamation process removes free water, most dissolved water, dissolved gas, particulate, carbon contamination, acids, oxidation by products, most color bodies, wear metals, and most additive chemicals.
TOP Oil Testers works on the development, production, market, technical and after-sale service for petrochemical analyzers, transformer substation electrical testing equipment and circuit detecting equipment. Main products are water content testers, surface/interfacial tension testers, flash point testers, kinematic viscosity testers, oil acidity testers, insulation oil dielectric strength testers, transformer characteristics testers, relay protection testers, high voltage testing equipment etc.