After-sales Service: | 1 Year |
---|---|
Warranty: | 1 Year |
Temperature Range: | Room Temperature ~500ºC |
Transport Package: | International Standard Export Case |
Trademark: | TOP TESTER |
Origin: | Chongqing |
Customization: |
---|
Suppliers with verified business licenses
The differential scanning calorimeter is a touch screen type, which can perform glass transition temperature test, phase transition test, melting and enthalpy value test, product stability, oxidation induction period test.
Standards compliant :
GB/T 19466.2- 2004/ISO 11357-2: 1999 Part 2: Determination of glass transition temperature
GB/T 19466.3- 2004/ISO 11357-3: 1999 Part 3: Determination of melting and crystallization temperature and enthalpy
GB/T 19466.6-2009/ISO 11357-6: 1999 Part 6: Determination of Oxidation Induction Time and Oxidation Induction Temperature
Industrial-grade 8-inch touch screen with rich display information.
Large LCD touch screen. Chinese and English interchangeable menu prompts; beautiful appearance and small size.
The detachable built-in furnace body design and modular design improve the installation process, and users can replace the furnace body by themselves.
USB communication interface, strong versatility, reliable and uninterrupted communication, supports self-recovery connection function.
Digital flowmeter, more accurate display.
Automatic switching of two air flow rates, fast switching speed and short stabilization time. At the same time, add a protective gas input.
The software is simple and easy to operate.
1. Temperature range: room temperature ~500ºC
2. Temperature resolution: 0.01ºC
3. Temperature fluctuation: ± 0. 1 ºC
4. Temperature repeatability: ± 0. 1 ºC
5. Heating rate: 0. 1 ~ 100ºC/min
6. Constant temperature time: recommended < 24h
7. Temperature control method: heating, constant temperature (automatic program control)
8. DSC range: 0 ~ ± 600mW
9. DSC Resolution: 0.01mW
10. DSC sensitivity: 0.01mW
11. Working power supply: AC220V/50Hz or customized
12. Atmosphere control gas: nitrogen, oxygen (automatically switched by the instrument)
13. Gas flow rate: 0-300m L/min
14. Gas pressure: 0.2MPa
15. Display mode: 24bit color, 8-inch LCD touch screen display
16. Data interface: standard USB interface
Suppliers with verified business licenses