Introduction
Portable octane analyzers are often used for on-site analysis of octane number of power gasoline, corresponding to the motor method and research method (RON and MON), and also suitable for cetane number analysis of diesel. In addition, the temperature of the analyzed liquid and the freezing point of diesel can also be measured, and the measurement results are displayed on the LCD liquid crystal display. This octane analyzer is widely used all over the world. Its measurement method conforms to international standards: octane measurement conforms to: ASTM D 2699-86, ASTM D 2700-86. Diesel cetane measurement conforms to: ASTM D 4737- 03, ASTM D 613, EN ISO 5165.
Measuring principle The principle of the TP-150 octane cetane number tester lies in the analysis of the insulation permeability of the octane number of gasoline and the cetane number of diesel oil and the charge characteristics of electromagnetic induction. By measuring the dielectric properties of the sample and comparing with the known parameters in the memory, the result is determined. The instrument is very sensitive and can measure small changes in dielectric parameters. It can detect octane number, cetane number and other petroleum product parameters. Functions Test octane number of automotive gasoline, as per ASTM D 2699-86, ASTM D 2700-86. Test cetane number of automotive gasoline as per ASTM D 4737-03, ASTM D 613, EN ISO 5165. Determination of diesel freezing point. Pour point depressant content for diesel fuel. Kerosene content of diesel fuel. Gasoline failure period (oxidation resistance), ASTM D 525.
The instrument can determine the
following parameters:
• Gasoline octane level in compliance with ASTM D 2699-86, ASTM D 2700-86.
Diesel fuel cetane level in compliance with ASTM D 4737-03, ASTM D 613, EN ISO 5165.
Reference parameters:
Diesel fuel type and сold filter plugging point (СFPP).
Gasoline breakdown time (oxidation stability) .
Kerosene percentage in diesel fuel.
Technical specifications:
Range of measured gasoline octane level - 40-135 ON
Max. permissible error in octane number measurement
- ±0.5 ON
Max. permissible discrepancy between parallel octane number measurements - ±0.2 ON
Range of measured gasoline breakdown time - 50-2400 min
Max. permissible error in gasoline breakdown time measurement - 10 min
Range of measured cetane level - 20-100 CN
Max. permissible error in cetane number measurement
- ±1.0 CN
Max. permissible discrepancy between parallel cetane number measurements - ± 0.5 CN
Max. permissible error in diesel СFPP measurement - ± 5 ºC
Range of measured kerosene percentage in diesel - 0- 95 %
Max. permissible error in kerosene percentage measurement - 3%
Measurement time - 1-5 sec
Threshold value for insufficient power supply alarm - 5.4 V
Instrument useful life - minimum 6 years
Overall dimensions of:
electronic module - 211х100х45 mm
sensor #1 - 60х100 mm
Instrument and sensor weight - 680 g
Package size 25x19x13 cm Package weight net - 1.0 kg


Production Lead-Time The first partial despatch of protective suits can be arranged for 3 days after payment received. Product can be delivered weekly depending on the quantity demanded.