1) Dimensions: 565 × 505 × 500mm]; (length × height × width) 2) Packing box dimensions: 730 × 650 × 610 [mm]; (length × height × width) 3) Column spacing size: 150mm 4) Column: two dedicated chromatography column 5) Instrument weight: 65Kg |
Temperature control index
|
Oven temperature control: 1) Room temperature plus 6 ºC ~ 399 ºC; (in increments of 1 ºC), (set effectively parameter up to 399 ºC, allowing using but not to guarantee the technical index) 2) Temperature fluctuation: no more than ± 0.1 ºC; (ambient temperature change 10 ºC or power supply voltage change 10%). 3) Temperature gradient: ± 1% (temperature range 100 ºC ~ 350 ºC). 4) Program heating step: 7 step 5) Heating rate: 0 ~ 40 ºC / min (set in increments of 0.1 ºC). 6) Cooling rate: not more than 3 min when oven cool from 200 ºC to 100 ºC 7) Time setting: 6000 (min) Thermal conductivity detector temperature control: Room temperature plus 20 ºC ~ 350 ºC (index parameters). (Set parameters: room temperature plus 6 ºC ~ 399 ºC effective, allowing using but not to guarantee the technical index ). Temperature control accuracy: not more than ± 0.1 ºC. The remaining heating zone temperature: Room temperature plus 20 ºC ~ 350 ºC (index parameters). (Set parameters: up to 399 ºC effective, allowing using but not to guarantee the technical index ). Temperature control accuracy: not more than ± 0.2 ºC. |
FID detector
|
Detection limit:≤8×10-12g/s(n-C16H34) Noise: ≤ 2 × 10-14A(0.02mv) Drift: ≤ 1.5 × 10-13A / 30min(0.15mv/30min) Dynamic linear range: ≥ 107 |
TCD detector
|
Sensitivity: ≤5000mv.ml/mg(n-C16H34) Noise: ≤2×10-14A (0.02mv) Drift: ≤1.5×10-13A/30min (0.15mv/30min) Dynamic linear range: ≥ 104 |
Application Environment
|
Ambient temperature: (5 ~ 35) ºC Relative humidity: no more than 85% Power Supply voltage: 220 ± 22V Power supply frequency: 50 ± 0.5Hz. Maximum power consumption: 2500W No Indoor corrosive gases. No strong mechanical vibration on table. No strong electromagnetic interference nearby. No dramatic indoor temperature change. No air convection |