TP-3600F Flame Atomic Absorption Photometer
I. Technical Requirements:
1 Adopt the latest multi-core embedded technology for fully automatic control and data processing of the instrument to ensure reliability.
2 High-performance LCD display, display parameters and measurement data.
3 Automatic adjustment of negative high voltage and lamp current.
4 Automatic conversion of spectral bandwidth, 0.1, 0.2, 0.4, 1.0, 2.0nm (five gears are optional).
5 Automatic control of wavelength scanning, automatic peak-seeking, international level of full-band rapid scanning and positioning, completed within 30 seconds.
6 Automatic adjustment of negative high pressure, lamp current; automatic setting of gas flow, automatic selection of the best combustion ratio for elemental analysis. Advanced and reliable safety protection system to protect the safety of operators in all aspects.
7 Automatic zero adjustment, can deduct the impact of zero point drift on data; automatic energy adjustment.
8 Automatic ignition, automatic flow setting, automatic protection for flame out, all-round gas leakage protection warning device, so that you can work with confidence and ease.
9 The best flame height and atomizer position can be selected to choose the best analysis conditions.
10 Excellent quality optical system, optical system suspension design, vibration, ambient temperature changes do not affect the performance of the instrument. Strong light energy, good stability.
11 Fast six-lamp automatic turret and wavelength scanning technology, available high-performance cathode lamp.
Technical Parameters
1 Optical system
1.1 Wavelength range: 190nm-900nm
1.2 Monochromator: C-T type monochromator with elimination of parallax, 1800 bars/mm
1.3, Spectral bandwidth: 0.1, 0.2, 0.4, 1.0, 2.0nm (five gears optional)
★1.4 Wavelength accuracy: ±0.1nm
★1.5 Wavelength repeatability:0.1nm
1.6 Baseline drift:0.004A/30min
2 Flame atomization system
2.1 Characteristic concentration:0.02.0ug/ml/1%
★2.2 Detection limit:( Cu ) 0.002μg/ml
2.3 Burner:100mm metal titanium burner, air-cooled premixed type
★2.4 Repeatability (RSD): ≤1%
2.5 Nebulizer: high-efficiency glass, all-plastic, stainless steel nebulizer
2.6 Atomization chamber: corrosion-resistant material atomization chamber
2.7 Position adjustment: the best height of the flame burner and the front and rear position automatically set adjustable, one minute to complete the flame / hydride change
2.8 Safety protection: with a variety of automatic safety protection functions, acetylene leakage alarm, shutdown system
Configuration list
No. |
Product Name |
Unit |
Qty |
Note |
1 |
Host |
set |
1 |
|
2 |
Air Compressor |
set |
1 |
|
3 |
Operating Software |
set |
1 |
Instructions included |
4 |
Hollow Cathode Lamp |
pcs |
3 |
Elements of customer's choice |
5 |
Nebulizer |
pcs |
2 |
1 installed, 1 spare part |
6 |
Screwdriver |
pc |
1 |
|
7 |
Spare parts box |
pc |
1 |
|
8 |
Air duct |
m |
5 |
Φ6mm |
9 |
Acetylene tube |
m |
5 |
Φ6mm |
10 |
Waste liquid pipe |
m |
1.5 |
Φ10mm pipe |
11 |
Three-core power cord |
pc |
1 |
|
12 |
Dust Cover |
pc |
1 |
|
13 |
Host signal cable |
pc |
1 |
USB cable |
14 |
Acetylene tube clamps |
pcs |
3 |
|
15 |
Acetylene gas source pipe |
m |
3 |
For acetylene cylinders |
16 |
3A Fuse |
pc |
1 |
|
17 |
1.5A fuse |
pc |
1 |
|
18 |
Pressure Reducing Valve |
pcs |
2 |
|
19 |
Acetylene+Bottle |
pc |
1 |
Customers prepared themselves |
About us
Hello Friend!
Chongqing TOP is a professional leading manufacturer and supplier of oil purifiers and various related testers. We consists of departments respectively for technological development, machinery manufacture, instrument manufacture, quality control etc. On basis of mature development and manufacture technology and highly qualified production management team, the products own high quality, easy operation, excellent performance and lone service life.
Customer satisfaction is our utmost priority thus, customer service team is always on the brink to help our customers with their queries at any time.
We believe in working as a unit with an attitude of complete trust and understanding and are determined to provide safe, reliable and competitively priced products to our customers.
TOP Oil Purifiers uses single or multi-stage filtration to remove, particulate, free water, and small quantities of dissolved water if necessary. The systems are usually customized to suite the specific customer application. Oil "regeneration", "reclamation", "recycling" are all terms used to describe the process of restoring used, aged, contaminated oils to "like new" reusable condition. The reclamation process removes free water, most dissolved water, dissolved gas, particulate, carbon contamination, acids, oxidation by products, most color bodies, wear metals, and most additive chemicals.
TOP Oil Testers works on the development, production, market, technical and after-sale service for petrochemical analyzers, transformer substation electrical testing equipment and circuit detecting equipment. Main products are water content testers, surface/interfacial tension testers, flash point testers, kinematic viscosity testers, oil acidity testers, insulation oil dielectric strength testers, transformer characteristics testers, relay protection testers, high voltage testing equipment etc.