Customization: | Available |
---|---|
After-sales Service: | 1 Year |
Warranty: | 1 Year |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
This instrument is designed and manufactured according to the requirements stipulated in the industry standard SH/T0165-92 / of the People's Republic of China "High Vacuum Distillation Determination Method for Petroleum Products in High Boiling Point Range". (equivalent ASTM D1160) .It is suitable for measuring the distillation range of petroleum products in high boiling point range such as wax oil and lubricating oil.
1. The instrument is composed of distillation flask heating furnace, heating power regulator, receiver measuring cylinder receiver and its air bath temperature control system, vacuum system, vacuum pressure gauge, lighting system, heating furnace cooling device and other parts.
2. Adopt high-precision U-shaped tube pressure gauge
3. High-quality vacuum pump is adopted, and the vacuum extraction is fast, which meets the test requirements of SH/TO165-92 "High Vacuum Distillation Method for Petroleum Products in High Boiling Point Range".
4. It adopts quartz tube electric heating furnace, which is durable and has a long service life. The heating power is stepless and continuously adjustable, and it is easy to use.
5. The air bath has a constant temperature and is equipped with a lighting lamp, which makes it easy to observe the refraction and the reading is accurate.
1. Working power supply: | AC 220V, 50Hz;1 Phase |
2. Heating power: | Distillation flask heater: 1200W; Receiver measuring cylinder heater: 400W; |
3. Vacuum measurement range: | 0~775mmHg±0.2mmHg(0kPa~103.3kPa±0.1kPa); |
4. Distillation flask heating furnace power: | 0 ~ 1000 W continuously adjustable; |
5. Air bath temperature control point of receiver measuring cylinder: | Continuously adjustable from room temperature to 90°C; |
6. Air bath temperature sensor: | Pt100 platinum resistance; |
7. Air bath temperature control method: | Automatic temperature control with digital display temperature controller; |
8. Temperature control accuracy: | Set temperature ±1°C; |
9. The maximum residual pressure of the vacuum pump: | ≤2mmHg; |
10. Instrument working conditions: | Ambient temperature: 5~35ºC; Relative humidity: < 85%. |