Constant Temperature Fast oil density, API degree tester
DST-300API-T, DST-120API-T
This series of products are mainly aimed at measuring the density and Baume degree of various chemical solutions. The constant temperature system and the density measurement system are integrated into one, and the liquid can be directly thermostatted to 20 °C; or at other specified temperatures, the rapid measurement of liquid density and API degree can be achieved. The constant temperature range of this instrument is 10°C~50°C, and the accuracy is 0.1°C; the whole machine has the characteristics of convenient operation, high precision and good durability; it is suitable for applications in quality management, formula development, cost control, etc.
Mainly for:
Petroleum industry, energy industry; suitable for ASTM D1298 standard specification, automatic display of density value and API degree.
Features:
- Can be constant temperature to 20°C, or other temperature, and can be set freely
- Constant temperature range: 10°C~50°C, accuracy 0.1°C
- Display density, API degree directly
- Adopt density measuring sensor imported from Germany and constant temperature system imported from Japan
- No cumbersome operation, easy to install
- With temperature compensation setting, air buoyancy density compensation setting
- Can be adapted to workshops, laboratories and other places
Operation:
- oIn the state of displaying 0.000, put the measuring cup containing the sample in the constant temperature bath.
- After the constant temperature reaches the specified temperature, hang the standard weight immersed in the sample on the center of the measuring rack with a hook, press the ENTER key, the display is the API value at the current temperature, and the symbol points to S2.
Technical indicators:
Model |
DST-300API |
DST-120API |
API precision |
0.1°API |
0.01°API |
AIP range |
0.1~100.0°API |
0.01~100.00°API |
Density accuracy |
0.001 g/cm3 |
0.0001 g/cm3 |
Density range |
0.001~99.999g/cm3 |
0.0001~99.9999g/cm3 |
temperature accuracy |
0.1°C |
0.1°C |
temperature range |
10°C~50°C |
10°C~50°C |
Measurement principle |
Archimedes principle |
Results display |
Density, API degree |
caliberation method |
One-button automatic calibration, automatic detection function |
How to check the measurement result |
Distilled water test |
output method |
RS-232C standard communication interface, convenient for test data output and printing |
Standard accessories |
Host, 100G weight, One power cord, 2 stainless steel hooks, 1 standard glass weight, 2 aluminum cups, 2 droppers |
Optional accessories |
DE-40 printer DE-20B anti-corrosive liquid measuring components |